Device Specifications:
Inlet Connection | Pin Index ISO 407 |
---|---|
Outlet Connection | G1/4 (1/4 BSP) |
Inlet Pressure | 1.5 to 20 MPa (20 to 200 bar) |
Standard Discharge (Q1) | 40 sl/min |
Maximum flow | 80 l/min |
Inlet filtration | 40 µm |
Over pressure valve set point | 700 kPa |
Service Interval | 4 years |
Intended Life | 12 years |
Regulatory | EC: MDD Class IIb Medical Device |
This device is designed for use with oxygen. It has a Pin Index inlet and a BS 5682 outlet.