Device Specifications:
Inlet Connection | CGA V-1 no. 660 |
---|---|
Outlet Connection | Rectus 20KA male |
Inlet Pressure | 2 to 20 MPa (20 to 200 bar) |
Inlet filtration | 40 µm |
Maximum flow | 12 L/min |
Over pressure valve set point | 700 kPa |
Service Interval | 2 years |
Intended Life | 12 years |
Applied Standards | EN ISO 10524-1 DD CEN/TS 14507-1 DD CEN/TS 14507-2 EN 837-1 |
Regulatory | MDD class IIb medical device |
Nitric oxide pressure regulators are a robust and reliable means to reduce the pressure of dilute nitric oxide gas mixtures from cylinder pressure to 4 bar (60 psi) for supply to inhaled nitric oxide (iNO) delivery systems.
Benefiting from our proven and trusted medical pressure regulator design, the precisely engineered nitric oxide pressure regulator provides a clean, safe, stable and consistent gas delivery.