Device Specifications:
Gas Compatibility | Oxygen and oxygen enriched air |
---|---|
Maximum Operating Pressure | 600 kPa*
*Maximum static pressure delivered by the gas supply source |
Flow Rate | 0.25* - 25 l/min
*The Firesafe Cannula Valve can be safely used at lower flow rates but may not meet the internal leak rate of 10 ml/min when activated in the event of a fire. |
Resistance to Flow | Per Firesafe Cannula valve:
≤ 0.75 kPa at 5 l/min
≤ 2.1 kPa at 10 l/min
≤ 4.2 kPa at 15 l/min
≤ 9.8 kPa at 25 l/min |
Intended Life | 4 years after date of manufacture (engraved on device) |
Application | Single patient use |
Packaging | 20 Firesafe oxygen therapy kits and 2 oxygen mask kits per box |
Connections | EN 13544-2, 6 mm nominal tubing connection |
Environmental Limits for Transport and Storage | Temperature: -20 °C to 50 °C
Humidity: 0 - 100 % RH |
Environmental Limits for Use | Temperature: 5 °C to 40 °C
Humidity: 0 - 100 % RH, including internal condensation |
Applied Standards | BS EN 13544-2:2002 Respiratory Therapy Equipment – Part 2: Tubing and connectors. |
Regulatory | EC: MDD 93/42/EEC Article 12, procedure pack |
Each pack consists of:
20 pre-assembled Firesafe oxygen therapy kits comprising:
Adult nasal cannula
10 meter long oxygen tubing
1 Firesafe Cannula valve
2 oxygen mask kits each comprising:
A medium concentration adult oxygen mask
Extension tubing