Firesafe Cannula Valve (100)

Part number: 827-2001x100

FireSafe Cannula Valve (100)

Double click on above image to view full picture

Zoom Out
Zoom In

More Views

  • FireSafe Cannula Valve (100)
  • FireSafe Cannula Valve (100)

Device Specifications:

Gas Compatibility Oxygen and oxygen enriched air
Maximum Operating Pressure

600 kPa*

 

*Maximum static pressure delivered by the gas supply source 

Flow Rate

0.25* - 25 l/min

 

The Firesafe Cannula Valve can be safely used at lower flow rates but may not meet the internal leak rate of 10 ml/min when activated in the event of a fire.

Resistance to Flow

≤ 0.07 kPa at 2 l/min

≤ 0.3 kPa at 5 l/min

≤ 1 kPa at 10 l/min

Intended Life 5 years, expiry date engraved on device
Application Single patient multiple use
Packaging 100 units per box (5 bags per box, 20 units per bag)
Connections EN 13544-2, 6 mm nominal tubing connection
Environmental Limits for Transport and Storage Temperature: -25 °C to 70 °C (-4 °F to 140 °F) Humidity: 0 - 100 % RH
Environmental Limits for Use Temperature: 0 °C to 50 °C (32 °F to 122 °F) Humidity: 0 - 100 % RH
Applied Standards EN 13544-2:2002 - Respiratory Therapy Equipment - Part 2: Tubing and connectors.
Regulatory EC: MDD Class IIa Medical Device

USA: FDA Registered Class 1 Medical Device; 510(k) exempt

GMDN Code and Term: 60391 - Fire safety valve

The Firesafe Cannula Valve is a thermal fuse designed to stop the flow of gas in the event that the downstream oxygen tubing is ignited.